Power amplifier (CE) |
|
Power
|
20Watt(20V@1A)
|
Compliance voltage
|
±20V |
Max. current |
±1A |
Control speed selection |
4 ea |
Bandwidth |
1MHz |
Slew rate |
8V/usec |
Potentiostat mode(voltage control) |
|
Voltage control |
|
- Control voltage range |
±10, ±1V, ±100mV |
- Voltage resolution |
16 bit per each range |
- Voltage accuracy |
±0.02% f.s(gain x1) |
- Max. scan range |
±10V vs. ref. E |
Current measurements |
|
- Current measurement range |
10 ranges (auto/manual setting)
10nA ~ 1A (1nA with gain) |
- Current measurement resolution |
16 bit; 30uA, 3uA, 300nA, 30nA, 3nA, 300pA 30pA, 3pA, 300fA, 30fA |
- Current measurement accuracy |
±0.03% f.s.(gain x1)>100nA |
Galvanostat mode(current control) |
|
Current control |
|
- Current control range
|
max. ±1A ± full sacle depending on selected range
|
- Current control resolution
|
16 bit; 30uA, 3uA, 300nA, 30nA, 3nA, 300pA,30pA, 3pA, 300fA, 30fA
|
- Current control accuracy
|
±0.03% f.s.(gain x1)>100nA |
Voltage measurements |
|
- Voltage measurement range
|
10V, 1V, 100mV |
- Voltage measurement resolution
|
16 bit; 0.3mV, 30uV, 3uV
|
- Voltage measurement accuracy
|
±0.02% f.s(gain x1) |
Electrometer |
|
Max. input voltage
|
±10V |
Input impedance
|
2x1013Ω||4.5pF
|
Bandwidth
|
>22MHz
|
CMRR
|
>114dB
|
EIS(Internal FRA) for system |
|
Frequency range
|
10uHz ~ 1MHz
|
Frequency accuracy
|
<0.01%>
|
Frequency resolution
|
5000/decade
|
Amplitude
|
0.1mV ~ 5V rms (potentiostatic) 0.1 ~ 70% f.s. (galvanostatic)
|
Mode
|
Static EIS: potentiostatic, galvanostatic, pseudo-galvanostatic, OCP Dynamic EIS: potentiodynamic, galvanodynamic Fixed frequency impedance: potentiostatic, galvanostatic, potentiodynamic, galvanodynamic Multisine EIS: potentiostatic, galvanostatic Intermittent PEIS/GEIS
|
Main system |
|
PCcommunication
|
USB2.0 high speed
|
Line voltage
|
100~240VAC, 50/60Hz
|
Size/Weight
|
411x321x165mm(WxDxH)
|
Max. output power
|
20Watt
|
Software |
|
Max. steps per experiments
|
1000
|
Shutdown safety limits |
voltage, current, power, temperature, etc. |
Max. sampling rate |
2usec or 3usec depending on data point number |
Min. sampling time |
unlimited |
Sampling condition |
time, dv/dt, dI/dt, temperature, etc.
|
System
|
|
Cell cable
|
1 meter shielded type (standard) working, reference, counter, working sense
|
Control
|
DSP with FPGA
|
DAC
|
2x16bit DAC(50MHz) for bias & scan 1X16bit DAC(1MHz) for analog output
|
Data acquisition
|
2x16bit ADCs(500kHz) for voltage, current
|
DAC
|
4x16bit ADCs(250kHz) for auxiliary voltage and temperature reading
|
Calibration
|
Automatic
|
Filter selection
|
4ea (5Hz, 1kHz, 500kHz, 5MHz)
|
Scan rate
|
0~200V/sec in common mode 0~5000V/sec in fast mode
|
Tablet PC
|
Included
|
Internal data memory
|
542,000 points
|
PC requirement
|
|
Operating system
|
Windows 7/8/10(32bit/64bit OS)
|
PC specification(minimum)
|
Pentium4, RAM 1GB or higher
|
Display
|
1600x900 high color or higher
|
USB
|
high speed 2.0
|
General
|
|
Dummy cell
|
One external dummy cell included
|
Thermocouple
|
K-type, 1.5 meter long(option)
|
Auxiliary cable
|
Option
|
Misc. cable
|
Option
|
Impedance analysis software |
ZMANTM software
|
DC data analysis software
|
IVMANTM software package
|